The EMOGA-5 can measure parameters such as methyl mercaptan, methyl sulfide, dimethyl disulfide, styrene, carbon disulfide, hydrogen sulfide, ammonia, trimethylamine, temperature, pressure, flow rate, humidity, etc., of which methyl mercaptan, methyl sulfide, dimethyl disulfide, styrene, carbon disulfide, hydrogen sulfide, ammonia and trimethylamine are measured by the characteristic absorption spectroscopy method; temperature by the platinum resistor (PT100) or the thermocouple, pressure by the pressure sensor and velocity by the Pitot differential pressure method, and humidity by the high temperature capacitance method. The characteristic absorption spectroscopy method can also measure benzene and toluene
Can directly monitor concentration values of each malodor gas component with fast response speed;
High-resolution spectrometer, with low detection limit;
PLS algorithm, which can effectively eliminate cross-interference between components;
Long-life pulsed xenon light source; no optical moving parts, high stability;
Standard CEMS, can be used with all kinds of systems.
petrochemical plants;wastewater treatment plants; waste incineration;coal chemical industry;Natural gas purification process gas analysis (trace H2 S); food processing, animal husbandry, catering and other occasions;living areas.
Repeatability | ≤3% |
Indication error | ±5%F.S. |
Zero drift | ±5%F.S./24h |
Span drift | ±5%F.S./24h |
Response time | ≤300s |
Data transmission protocol | national standard 212, Mudbus, local standard protocols Communication interface: RS232, USB&TCP |
Power supply | AC (220±22)V; (50±1)Hz |
Rated power | 2500W |
Ambient temperature | 15℃~35℃ (indoor); -20℃~50℃ (outdoor) |
Ambient humidity | ≤ 85%RH, non-condensing |
Heat tracing temperature | 120ºC~150ºC |
Flow rate | (2±0.5)L/min |